GASTEC CORPORATION - For all types of gas and vapour

ANASYS

Operation of each logging conditions and treatment of the measured value of the instrument

I will explain the measurement interval of the instrument, handling of measurement by the sensor full scale, the differences between logging operation by logging conditions.

Measurement interval

This is the timing to update the value of body liquid crystal display by obtaining the gas concentration is measured. Once a second in the state where no logging in the instrument, I'm going once measured two seconds while logging. The reason for the difference of two seconds and this one second has been set at one second shorter interval for performing manually span calibration and zeroing in the state in which two seconds, not logging longer interval to reduce current consumption in logging in .

Handling of the value measured by the sensor full scale

Showed how are treated measurements by the sensor full scale 100ppm, 500ppm, 1000ppm of when the gas concentration was varied as shown in the graph below as an example. Measurements at each full scale is shown in Figure 1 below in the table below. 1000ppm sensor has a measurement value of 0ppm to no more than 10ppm of minimum resolution between 0-26 seconds elapsed time. Minimum resolution smaller values are treated rounded to a lower value side in this way.

GHS8AT_FS_MEASURE
Figure1. The measurement results due to the difference of the sensor full scale of the instrument and gas concentration

Difference in the resolution by the sensor full scale

Difference in the minimum resolution of the sensor full scale is in Table below.

Table1. Relationship of the minimum resolution and sensor full scale
Full scale[ppm] Sensor Model minimum resolution[ppm]
10 H2S-520E 0.1
100 H2S-521E 1
500 H2S-522E 1
1000 H2S-523E 10
3000 H2S-524E 25

Operation of each logging conditions

There are four types of real (REAL), minimum (MIN), maximum (MAX), average (AVE) logging conditions. Shows the measurement results for each logging conditions if you change gas concentration is as shown in the graph in Figure 2 below as an example. In this example, I have one minute the sensor full scale 100ppm logging interval.

GHS8AT_LOG_CONDITION
Figure2. Difference of logging data by logging conditions

The information about the data calculated in the case where the average logging conditions

Average values ​​are calculated the average value of all the data in the logging interval. Measurement interval because it is two seconds, for example, we have calculated the average of 150 data in the case of 5 minute logging, the average of 30 data inteh case of 1 minute logging. In the case of Figure 2, we have divided after 30 to the sum of the measured data of 30 to 60 seconds from 2 seconds elapsed time. It becomes 8.3667 Dividing by 30 this total of up to 60 seconds from the second becomes the 251, but will be saved as a truncated 8 [ppm] a decimal point as logging data. Truncated value of the average value at the time of calculation will be rounded down to the decimal point all of the full-scale other than 10ppm as the following table by each sensor full scale.

Table2. Handling of the average value calculation results logging conditions average
Full scale[ppm] Sensor Model Average value calculated data handling Minimum value[ppm]
10 H2S-520E Second decimal place truncated 0.1
100 H2S-521E Decimal point truncated 1
500 H2S-522E Decimal point truncated 1
1000 H2S-523E Decimal point truncated 1
3000 H2S-524E Decimal point truncated 1